Wide Area Oxygen Transmitter (CI-PC986)
-
Meets measurement accuracy requirements across a wide range of 1 ppm to 25%.
-
The transmitter features voltage regulation and automatic flow control, making it suitable for positive pressure applications.
-
Uses a digital sensor that allows on-site replacement without requiring recalibration.
-
Includes an overvoltage (pressure) protection function: when pressure in the test chamber exceeds the preset range, the system automatically shuts off the gas path to prevent sensor damage caused by excessive or insufficient pressure.
Principle: Electrochemical
Measuring range: 0~10/100/1000ppm/1/25.00% O2
Accuracy:
0.01~9.99ppm O2: ±5%FS
10.0~99.9ppm O2: ±3%FS
100~1000ppm O2: ±2%FS
Repeatability:
0.01~9.99ppm O2: ±2.5%FS
10.0~99.9ppm O2: ±1.5%FS
100~1000ppm O2: ±1%FS
Stability:
0.01~9.99ppm O2: ±2.5%FS/7d
10.0~99.9ppm O2: ±1.5%FS/7d
100~1000ppm O2: ±1%FS/7d
Response time: T90<60s (25°C)
Communication: RS485
Analog Signal: 4~20mA (standard)/ 0~20mA (optional)