CI-PC35 Online Wall-Mounted Dew Point Analyzer
Microprocessor as the core, good long-term stability, high reliability;
Using ultra-thin alumina capacitive sensor as measuring unit, high precision, long calibration cycle;
Dew point value and ppm are displayed at the same time to meet the different needs of users;
Menu lock switch: prevent misoperation to change the analyzer parameters affect its normal operation;
An alarm will be issued when the dew point value deviates rom the alarm value, which can be set arbitrarily in the full scale range.
Principle: Ultra-thin Alumina Capacitive
Measuring Range:
-60°C~+20°C (basic)
-80°C~+20°C (standard)
-100°C~+20°C (extension)
Linear Error: ±3°C(-80~-60°C), ±2°C(-60~+20°C)
Repeatability: ±0.5°C
Zero Drift: ±1%FS/7d
Range Drift: ±1%FS/7d
Response Time: 63%[90%] range at +20°C and 1bar pressure
-60→-20°C Td 5s[10s]
-20→-60°C Td 45s[10min]
Power Supply: DC 24V, 1A
Communication: RS485(standard)/RS232(optional)
Ambient Temperature.: 0~+45°C
Ambient Humidity:<90%RH
Analog Signal: 1 group, 4~20mA/0~20mA/0~1V/0~5V/0~10V/1~5V, can be set mode freely, Load:<500Ω
Sample Gas Pressure:Atmospheric Conditions±10% (Gas outlet should be atmospheric conditions)
Sample Gas Flow: 3~5L/min