Online Gas Analysis System (CI-XT60EO-O2X)

Home > Product > CI-XT60EO-O2X Online Gas Analysis System

Online Gas Analysis System (CI-XT60EO-O2X)

MODEL : CI-XT60EO-O2X

The CI-XT60EO-O2X is a chemical explosion-proof online oxygen analysis system equipped with a pretreatment unit independently developed by our company. Engineered with corrosion-resistant and explosion-proof technologies, it fully complies with the safety standards required for hazardous and explosive production environments. The integrated pretreatment system includes essential components such as a vacuum generator, automatic condensate drainage unit, filtration module, calibration valve, oxygen analyzer, and a robust explosion-proof cabinet. This comprehensive configuration ensures reliable, safe, and continuous oxygen monitoring in demanding industrial applications.

Feature:
High Precision & Fast Response: Delivers accurate oxygen measurements with rapid response time and excellent long-term stabilit;
Reliable & Low Maintenance: Designed for high reliability and extended calibration intervals, reducing maintenance efforts and costs;
Explosion-Proof Laser Technology: Equipped with one of the world’s most advanced explosion-proof laser oxygen analyzers, ensuring safe and precise performance in hazardous environments;
Advanced Laser Detection Principle: Utilizes a laser-based optical detector to accurately measure atmospheric oxygen concentration;
Optimized Air Circuit Design: Features a uniquely engineered air path with excellent sealing, impermeability, and resistance to moisture absorption, ensuring measurement integrity;
Interference Compensation: Built-in compensation function effectively eliminates the impact of interfering gases, enhancing measurement accuracy;
Analog Output: Provides a standard 4–20 mA analog output (1 channel) for seamless integration with control systems;

Performance Parameters:
Principle: Tunable Diode Laser Absorption Spectroscopy (TDLAS)
Measuring Range: 0~5.0% O2
Accuracy: ±0.2% O2
Stability: <±1% reading/year
Zero Drift: ±0.1% O2/year
Response Time: Lag Time: T10 ≤5s
Rise Time: T10~T90 ≤2s
Work Pressure: 0.8~1.4 bar
Storage Temperature: +10~+50°C
Analog Signal: 4~20mA
Analyzer Life: More than 5 years (normal use)
Working Conditions:
Power Supply: AC 220V 50Hz
Ambient Temperature: +15~+45°С
Relative Humidity: ≤85%RH
Sample Gas Temperature: Normal temperature (≤35°C)
Sample Gas Pressure: 30 КРа
Sample Gas Flow: ≤3L/min
Dust Concentration: Zero
Pressure Range: 0~0.2MPa
Nitrogen Purging: Requirements: No oil, no water, no dust
Pressure: ≥0.5Mpa
Compressed Air: 0.3~0.5Mpa
Air flow: ≤0.5m/s
Ambience: The installation environment must be free of explosive, corrosive, and other harmful gases, and should have adequate ventilation.
Safety Performance: The insulation resistance shall not be less than 5 MΩ. Under a test voltage of 1.25 kV, no breakdown or arcing shall occur within 1 minute.
Configuration And Output:
Reaction Speed: Lag time:T10 ≤30s
Rise time: T10~T90 ≤2s
Zero Drift: ≤±1% FS/7d
Range Drift: ≤±1% FS/7d
Load: ≤750Ω
Analog Signal: 4~20mA
Sampling Method: Ejection extraction
System Cabinet Parameters:
System Cabinet: Outdoor stainless steel rainproof cabinet
Specification: LxWxH= 500 × 400 × 1300 (mm)
Ex-mark: Ex dell CT6 Gb
Location Of Installation: Outdoor

Related products

CI-XT683 Online Biogas Analysis System
Online Coke Oven Gas Analysis System (CI-XT100B)
Online Coke Oven Cycle Monitoring System (CI-XT300)
Process Analysis System For Gas Generator (CI-XT6400)
Brazing Furnace Online Oxygen Analysis System (CI-XT6003C)
CO2 Recycling Online Monitor System (CI-XT800)
CI-XT62B Heat Treatment Furnace
Online Gas Analysis System (CI-XT60EO-O2X)
CI-XT6700 Online Calorific Value Blast Furnace Gas Analysis System
Online Oxygen Analysis System for Centrifuge (CI-XT60E3-O2D)
CI-XT5100 Fixed Pollution Source VOCs Online Monitoring System
CI-XT20-1 Desktop Medical Oxygen Quality Analyzer
CI-XT20-2-Q Online Medical Oxygen Quality Analysis System
Air Separation Process Online Analysis System (CI-XT64B)