Online High Temperature Humidity Analysis System (CI-PC196)
The probe is directly inserted into the high-temperature flue gas to measure the humidity and oxygen concentration in situ;
With automatic calibration function, it can automatically complete the calibration according to the cycle to improve the measurement accuracy;
With automatic backflush function, keep the probe clean, reduce user maintenance workload, and prolong the service life of the probe;
The probe has pressure monitoring to judge whether the sample gas collection is normal.
Principle: Double Cell Thick Film Zirconia
Measuring range: H2O 0~40.00% (configurable)
O2 0~25.00%
Accuracy: H2O ±2%FS
O2 ±2%FS
Repeatability: H₂O ±1%FS
O2 ±1%FS
Stability: H2O ±1%FS/7d
O2 ±1%FS/7d
Response Time: T90<60s (after full preheating)
Power Supply: AC 100~240V, 50/60Hz
Ambient Humidity: <80%RH
Storage: Temperature: -20~+60°C
Storage Humidity: <90%RH
Inlet pressure: Atmospheric Conditions ±5%
Sample Gas Flow: 1~2L/min (1.5L/min recommended)
Remarks: Ensure that there are minimal dust, no smoke, water vapor and corrosive gas